Maintenance and maintenance of light incubator

The light incubator has the function of over-temperature and abnormal sensor protection to ensure the safety of the instrument and samples; the optional full-spectrum plant growth lamp is conducive to plant growth and improves disease resistance. With power-off memory, power-off time automatic compensation function; constant temperature control system, quick response, high temperature control accuracy. Maintenance and maintenance of light incubator: 1. The incubator shell should be reliably grounded. 2. When cooling the incubator, it is not advisable to make the difference between the temperature in the chamber and the ambient temperature greater than 25 degrees. 3. When carrying, repairing and maintaining the incubator, avoid collision, shaking and vibration; the maximum inclination is less than 45 degrees. 4. In order to ensure the effective heat dissipation of the condenser, the distance between the condenser and the wall should be greater than 100mm. There should be a gap of 50mm on the side of the box and at least 300mm of space on the top of the box. 5. The incubator should be placed in a cool, dry, well-ventilated place away from heat sources and sunlight. Place it steadily to prevent vibration and noise. 6. The instrument does not work suddenly, please check whether the fuse tube (behind the box) is burnt out, and check the power supply.